Instruments for X-ray Analysis: X-ray tubes
Загрузить Adobe Flash Player

Instruments for X-ray Analysis: X-ray tubes

Печать PDF

   X-ray tubes "Renis-M"

Miniature X-ray tube TRM10/30

Anode current

0 - 350 microA

Anode voltage

10 - 30 kV

Maximum power

10 W

Filament voltage

1,36 V

Filament current

300 mA

Filament power

0,4 W

Stability

0.2% per 4 hour

Typical focal spot size

<200  micron

Insulating medium

Oil, compound

Cooling

Forcibly

Overall dimensions:

 

- diameter

15 mm

- length

49 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

17 gram

 

Miniature x-ray tube TRM 10/40

Anode current

0 - 250 microA

Anode voltage

10 - 40 kV

Maximum power

10 W

Filament voltage

1,36 V

Filament current

300 mA

Filament power

0.4 W

Stability

0.2% per 4 hour

Typical focal spot size

<200  micron

Insulating medium

Oil, compound

Cooling

Forcibly

Overall dimensions:


- diameter

15 mm

- length

59 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

20 gram

Small-sized X-ray tube TRM 30/40

 

Anode current

0.01 - 1 mА

Anode voltage

3 - 40 kV

Maximum power

30 W

Filament voltage

6.3 V

Filament current

330 mA

Filament power

2 W

Stability

0.2% per 4 hour

Typical focal spot size

<100  micron

Insulating medium

Oil, compound

Cooling

Forced

Overall dimensions:

 

- diameter

25 mm 

- length

88 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

105 gram

 

Small-sized X-ray tube TRM 50/50

 

Anode current

0,01 - 1 mA

Anode voltage

10 - 50 kV

Maximal power

50 W

Filament voltage

12 V

Filament current

70 mA

Filament power

0.8 W

Stability

0.2% per 4 hour

Focal spot size

<200 micron

Bias voltage

0 - (-200)V

Insulating medium

Oil, compound

Cooling

Forced

Overall dimensions:


- diameter

25 mm

- length

103 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

140 gram

 

Small-sized X-ray tube TRM 10/50

 

Anode current

0 - 200 microA

Anode voltage

5 - 50 kV

Maximal power

10 W

Filament voltage

12 V

Filament current

70 mA

Filament power

0.8 W

Stability

0.2 per 4 hour

Length of outrigger anode

60 mm

Diameter of outrigger anode

8 mm

Target type

Transmission

Bias voltage

0 - (-200)V

Insolating medium

Oil, compound

Cooling

Naturally

Overall dimensions:

 

                                - diameter

25 mm

                                - length 

130 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

105 gram

 

Small-sized X-ray tube TRM 10/50M

 

Anode current

0 - 200 microA

Anode voltage

5 - 50 kV

Maximal power

10 W

Filament voltage

12 V

Filament current

70 mA

Filament power

0.8 W

Stability

0.2% per 4 hours

Length of outrigger anode

165 mm

Diameter of outrigger anode

10 mm

Target type

Reflecting

Bias voltage

0 - (-200)V

Insulating medium

Oil, compound

Cooling

Forced

Overall dimensions:

 

- diameter

25 m

- length

264 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

210 gram

 

Small-sized X-ray tubes TRM 400/80

 

Anode current

0,01 - 7 mA

Anode voltage

10 - 80 kV

Maximal power

400 W

Filament voltage

6.3 V

Filament current

300 mA

Filament power

2 W

Stability

0.2% per 4 hours

Typical size of focal spot

<1,5  mm

Bias voltage

0 - (-200)V

Insulating medium

Oil, compound

Cooling

Water (3 l/ min)

Overall dimensions:

 

- diameter

44 mm

- length

130 mm

Target material

Cu, Mo, Ag, Cr et al.

Mass

350 gram